1. Combined effect of low-frequency high‑intensity ultrasound and streptokinase on the efficiency of thrombi destruction

I. Adzerikho1, A. Bubulis2, V. Jurėnas3, V. Minchenya4, V. Veikutis5, A. Vilkauskas6, A. Puodžiukynas7

1State Higher Educational Establishment “Belarusian Medical Academy of Post-Graduate Education”,
P. Brovki Str. 3, 220013 Minsk, Republic of Belarus

2, 3, 6Kaunas University of Technology, Donelaičio 73, LT-44029 Kaunas, Lithuania

4Belarusian National Technical University, 65 Independence Ave., 220027 Minsk, Republic of Belarus

5, 7Lithuanian University of Health Sciences, Institute of Cardiology, Sukilėlių 17, Kaunas, Lithuania

2Corresponding author

E-mail: 2algimantas.bubulis@ktu.lt, 3vytautas.jurenas@ktu.lt, 4vlad_minch@mail.ru

(Received 22 October 2012; accepted 4 March 2013)

Abstract. Results of numerous experimental studies indicate that low-intensity ultrasound (US) intensifies the drug induced fibrinolysis. However, exits opinion, what high-intensity ultrasound could be not recommending for use in combination with thrombolytic agents due to the transience of exposure of thrombus to ultrasound and relatively low speed of the enzymatic process. We also hypothesise that high-intensity low-frequency US can accelerate the drug induced fibrinolysis. During in vitro studies, we showed that the administration of streptokinase (SK) prevents the formation of particle conglomerates resulting due to destruction of fibrin clots by using low-frequency high-intensity US. In addition, it was established that in conditions of the combined effect of streptokinase and high-intensity US, the process of acoustic-mechanical destruction of thrombus is prevailing, and the fibrinolytic effect of streptokinase is generally witnessed after the termination of US and manifests itself in enzymatic lysis of both resultant particles and residual clot mass. In this study we verified the effect of low-frequency high-intensity ultrasound on the dynamics and efficiency of fibrinolysis induced by streptokinase on various structure thrombi model, characteristic features of the fibrin clot network structure occurring during the combined effect of ultrasound and streptokinase.

Keywords: low-frequency, ultrasound, piezoelectric transducer, drug induced fibrinolysis, combined thrombolytic therapy.

References

[1]        Lauer C. G., Burge R., Tang D. B., et al. Effect of ultrasound on tissue-type plasminogen activator-induced thrombolysis. Circulation, Vol. 86, No. 4, 1992, p. 1257‑1264.

[2]        Behrens S., Daffertshofer M., Spiegel D., et al. Low-frequency, low-intensity ultrasound accelerates thrombolysis through the skull. Ultrasound. Med. Biol., Vol. 25, No. 2, 1999, p. 269‑273.

[3]        Suchkova V. N., Baggs R. B., Francis C. W. Effect of 40-kHz ultrasound on acute thrombotic ichemia in a rabbit femoral artery thrombosis model. Enhancement of thrombolysis and improvement in a capillary muscle perfusion. Circulation, Vol. 101, No. 19, 2000, p. 2296‑2301.

[4]        Tachibana K. Enhancement of fibrinolysis with ultrasound energy. J. Vasc. Interv. Radiol., Vol. 3, No. 2, 1992, p. 299‑303.

[5]        Hong A. S., Chae J. S., Dubin S. B., et al. Ultrasonic clot disruption: an in vitro study. Am. Heart J., Vol. 120, No. 2, 1990, p. 418‑420.

[6]        Trubestein G., Engele C., Etzel F., et al. Thrombolysis by ultrasound. Clin. Sci. Mol. Med., Vol. 3, 1976, p. 697‑698.

[7]        Philippe F., Drobiski G., Bucherer C., et al. Effects of ultrasound energy on thrombi in vitro. Cath. Cardiovasc. Diagn., Vol. 28, No. 2, 1993, p. 173‑178.

[8]        Adzerikho I. E., Mrochek A. G., Dmitriev V. V., et al. Ultrasound fibrin clot destruction in vitro in the presence of fibrinolytic agent. Ultrason. Sonochem., No. 8, 2001, p. 315‑318.

[9]        Ernst A., Schenk E., Gracewski G., et al. Ability of high-intensity ultrasound to ablate human atherosclerotic plaques and minimize debris size. Am. J. Cardiol., Vol. 68, No. 2, 1991, p. 242‑246.

[10]     Bielychier V. A., Varietskhaya T. V., Vieriemieienkho K. N., et al. Methods of Determination of Fybrynogen and Components of Fybrynogen of the Plazma of Human Blood: Methodical Recommendations. Kiev, 1983, 20 p., (in Russian).

[11]     Goldstein J., Newbern D., Etchlin P., et al. Raster Electronic Microscopy and X-ray Microanalysis. In two Volumes, Moscow, Mir, 1984, Vol. 1: 303 p., Vol. 2: 348 p., (in Russian).

[12]     Ghlanch S. Medical – Biological Statistics. Moscow, Practica Publishing House, 1999, 459 p., (in Russian).

[13]     Lakin G. F. Biometry. Moscow, Higher School Publishing House, 1990, 352 p., (in Russian).

[14]     Sabovic M., Lijnen H. R., Keber D., Collen D. Correlation between progressive absorption of plasminogen to blood clots and their sensitivity to lysis. Thromb. Haemost., Vol. 64, No. 3, 1990, p. 450‑454.

[15]     Glover C. J., McIntire L. V., Leverett L. B., et al. Effect of shear stress on clot structure formation. Trans. Am. Soc. Artif. Int. Organs., Vol. 20, 1974, p. 463‑468.

[16]     Sobel B. E., Nachowiak D. A., Fry E. T. A., et al. Paradoxical attenuation of fibrinolysis attributable to “plasminogen steal” and its implications for coronary thrombolysis. Coronary Artery Dis., Vol. 1, 1990, p. 111‑119.

[17]     Carr M. E., Hardin C. L. Fibrin has larger pores when formed in the presence of erythrocytes. Am. J. Physiol., Vol. 253, No. 5, 1987, p. 1069‑1073.

[18]     Rosenschein U., Frimerman A., Laniado S., et al. Study of the mechanism of ultrasound angioplasty from human thrombi and bovine aorta. Am. J. Cardiol., Vol. 74, No. 12, 1994, p. 1263‑1266.

[19]     Anand S., Wu J. H., Diamond S. L. Enzyme mediated proteolysis of fibrous biopolimers: disolution front movement in fibrin or collagen under conditions of diffusion or connective transport. Biotechnol. Bioeng., Vol. 48, 1995, p. 89‑107.

[20]     Nyborg W. L. Ultrasonic microstreaming and related phenоmena. Br. J. Cancer., Vol. 45, No. 5, 1982, p. 156‑160.

[21]     Miller D. L., Thomas R. M., Williams A. R. Mechanisms for hemolysis by ultrasonic cavitation in the rotating expoure system. Ultrasound. Med. Biol., Vol. 17, No. 2, 1991, p. 171‑178.

[22]     Everbach E. C., Francis C. W. Cavitational mechanisms in ultrasound-accelerated thrombolysis at 1 MHz. Ultrasound. Med. Biol., Vol. 26, 2000, p. 1153‑1160.

[23]     Kiseliov M. G., Minchenya V. T., Stepanenko D. A. Ultrasound in Medicine. Minsk, “Technical Literature” Publishing House, 2009, 428 p., (in Russian).

[24]     Miller D. L. Particle gathering and microstreming near ultrasonically activated gas-filled micropores. J. Acoust. Soc. Am., Vol. 84, No. 4, 1988, p. 1378‑1387.

[25]     Alliger H. Ultrasonic disruption. Am. Laboratory, No. 7, 1975, p. 75‑85.